Print version
Multichannel Frequency Comparator VCH-315Frequency comparator VCH-315 is intended for precise phase and frequency instability measurements. It contains eight identical measuring channels and performs high precision phase and frequency comparison.
Key applications
- verification of metrological parameters of precision frequency signals sources;
- time and frequency etalon systems;
- computed control measustem;
- scientific research measurements.
Specifications
Input signals: 5 or 10 or 100 MHz nominal frequency, (0.8÷1.2) V into 50 Ω load.
Input impedance: 50 Ω.
Noise passband: 10 Hz.
Maximal measured relative frequency difference: ±5.0·10–9.
Averaging time range: from 1 s up to 106 s.
Metrological characteristics are given in the table:
Main measurement error * (Frequency instability, inserted by noises of measurement channels) |
Averaging time, (τ) |
Allan deviation noise floor |
1 s |
1.5·10–13 |
10 s |
2.0·10–14 |
100 s |
3.0·10–15 |
1000 s |
5.0·10–16 |
1 day |
1.0·10–16 |
Specified under condition: ambient temperature changing rate <1°Ñ/hour. |
Interface: RS-232; USB.
Application software
- runs under Microsoft Windows 2000,XP, Vista, 7;
- calculates phase changing and frequency difference for each signals pair, two-samples Allan deviation, N-samples Allan deviation;
- represents measurement results as tables and plots, stores information in files.
Operating temperature range: +5 °Ñ to +40 °Ñ.
Interface: RS-232; USB.
Power supply: AC(198÷242)V, (50÷60)Hz.
Power consumption: ≤40 V⋅A.
Dimensions (W×H×D): 483×133×370 mm.
Weight: 8 kg.
Warranty: 3 years.
Live time: 15 years.
Download manual

|
|