Passive Hydrogen Maser VCH-1006

VCH-1006 Passive Hydrogen Maser is a small size hydrogen maser with the best frequency stability. Extremely high frequency stability is provided by state-of-the-art technology. Full digital processing of modulation and servo loop signals is realized. Ideal for high-accuracy measurements applications.

Key applications:

– National Time Keeping Service.
– Space tracking and navigation.
– Verification of metrological parameters of frequency signals sources.
– Scientific research.

Specifications

Output signals:
Sine: 5 MHz; 10 MHz; 100 MHz, (1±0.2) V RMS into 50 Ω load.
Pulse: 2,048 MHz (square pulse), (1,5÷2,8) V (pp) into 75 Ω (ITU-T G.703 part13).
1Hz, positive polarity pulse, width (15±5) μs, TTL level at 50 Ω, rise time: <15 ns.

Metrological characteristics are given in the table:

Relative frequency accuracy ±3·10 –13 (factory calibration)
Output signals frequency corrector resolution
tuning range
1·10 −15
1·10 −10
Frequency stability (Allan deviation at 25±1C°, environmental effects are excluded) 1 s ≤7.0·10 −13
10 s ≤2.0·10 −13
100 s ≤7.0·10 −14
1 hour ≤2.0·10 −15
1 day ≤7.0·10 −15
Phase noise spectral density (dBc/Hz) (5 MHz output) Frequency offset Spectral density
1 Hz ≤−110
10 Hz ≤−130
100 Hz ≤−145
1000 Hz ≤−155
Manual synchronization to external 1 pps TTL signal accuracy ≤±50 ns
Temperature sensitivity in temperature operating range (+5÷+35 C°) ≤±2·10 −14 1/C°
Magnetic sensitivity ≤±2·10 −14 1/Oersted

Digital control and monitoring: all operating parameters on local LCD display or remotely.
Interface: RS-232C.
Power: AC, (100÷240)V, (50÷60)Hz, or DC, (27 +3 -5 )V.
Power consumption: 120 VA AC, 80 W DC.
Dimensions (W×H×D): 470 mm×200 mm×530 mm.
Weight: 31 kg.
Warranty: 3 years.