Frequency distribution amplifier is designed for ultra-low-noise distribution of sinusoidal signals 1–100 MHz with unity gain. It allows checking for signals availability via RS-232 interface. The basic model has 3 inputs / 12 outputs (3 modules 1 / 4). For the information about possible modifications, please contact us.
Key applications:
– time and frequency high precision measurement systems;
– time and frequency etalon systems.
Manual for VCH-605
– Operational Manual
download
Specifications
Input signals: sinusoidal 1–100 MHz.
Maximal power of the input signals: 17dBm.
Amplifier gain: (0±1) dB.
All characteristics presented below are guaranteed for sinusoidal signals 5±0.00005, 10±0.0001, 100±0.001 MHz and RMS input voltage 1±0.1 V at 50 Ohm
Isolation between channels no more, than the values in the table
Signal frequency | |||
---|---|---|---|
5 MHz | 10 MHz | 100 MHz | |
Output/Output | −120 dB | −120 dB | −100 dB |
Input/Output | −120 dB | −120 dB | −100 dB |
Input/Input | −120 dB | −120 dB | −100 dB |
Harmonic distortions introduced by the amplifier no more than –40 dB at 5, 10 MHz and no more –35 dB at 100 MHz.
Allan deviation 1 sec introduced by the amplifier is no more than 4.0·10 –14 .
Ultra low phase noise (single-side power spectral density) added by the amplifier is no more than the values in the table
Frequency offset | Signal frequency | ||
---|---|---|---|
5 MHz | 10 MHz | 100 MHz | |
1 Hz | −141 dBc/Hz | −141 dBc/Hz | −130 dBc/Hz |
10 Hz | −150 dBc/Hz | −150 dBc/Hz | −145 dBc/Hz |
100 Hz | −155 dBc/Hz | −155 dBc/Hz | −150 dBc/Hz |
1 kHz | −161 dBc/Hz | −161 dBc/Hz | −155 dBc/Hz |
10 kHz | −163 dBc/Hz | −163 dBc/Hz | −160 dBc/Hz |
Temperature coefficient no more than 5 ps/°С.
Power supply: AC (100÷240) V, (50÷60) Hz; DC (22÷30) V.
Power consumption is no more than 45 V·A (for AC) and no more than 40 W (for DC).
Status information is available via RS-232 interface.
Weight is no more than 4 kg.
Dimensions: 310.0×483.0×44.0 mm (1 U)