Frequency Calibrator VCH-313

Download PDF

Frequency calibrator VCH-313 is intended for calibration and verification of different frequency signal sources, using precision time scale signal, transmitted by satellite navigation systems GPS/GLONASS. The instrument can provides periodical frequency correction of the calibrating frequency standard in automatic mode.

Key applications
– Verification of metrological parameters of precision frequency signals sources;
– time keeping metrology systems;
– scientific research measurements.

Specifications

Input signals serviceable for frequency calibration:
– 5 or 10 MHz, sine, 1 V into 50 Ω load;
– 2,048 MHz, pulse, (ITU-T G.703); The frequency of input signal is recognized automatically.

Frequency calibration error depends on input signal frequency “F” and measurement time “T” and is determined by formula ΔF/F=1/FT. Values of calibration errors are given in the table:

Input frequency “F” Calibration error depends from measurement time “T”
1 hour 6 hour 24 hours
2,048 MHz ≤6.0·10 −10 ≤1.0·10 −10 ≤3.0·10 −11
5 MHz ≤1.5·10 −10 ≤2.5·10 −11 ≤6.0·10 −12
10 MHz ≤3.0·10 −11 ≤5.0·10 −12 ≤1.0·10 −12

The best frequency calibration accuracy (≤1.0·10 –13 ) is realizing for measurement time T=10 days.
Results of calibration in relative units are available on the display at the front panel of the indtrument and remotely through interface.
Interface: RS-232; USB.
Power supply: AC(198÷242)V, (50÷60)Hz; DC(38.4÷57.6)V.
Power consumption: ≤25 W.
Start up time: ≤25 min.
Dimensions (W×H×D): (without antenna and a cable) — 240×140×330 mm.
Antenna: dimensions 185×∅181 mm. Weight 0,5 kg. Antenna cable’s lenght is 60 m.
Weight: 4.5 kg.
Warranty: 3 years.
Live time: 15 years.