Passive Hydrogen Maser VCH-1009

Download PDF

The VCH-1009 Passive Hydrogen Maser is a compact hydrogen maser with excellent frequency stability and utilises state-of-the-art technology. Full digital processing of the modulation and servo loop signals makes it ideal for high accuracy applications. An internal GPS/GLONASS option is available and provides automatic output signal frequency calibration.


Key applications
– National Time Keeping Service;
– Space tracking and navigation;
– Verification of frequency signals;
– Scientific research.

Specifications

Output signals:
Sine: 5 MHz; 10 MHz; 100 MHz, (1±0.2) V RMS into 50 Ω load.
Pulse: 2.048 MHz (square pulse), 1.5÷2.8 V (pp) into 75 Ω (ITU-T G.703 part15).
1Hz, positive polarity pulse, width 15±5 μs, TTL level at 50 Ω, rise time: <2 ns.

Metrological characteristics are given in the table:

Frequency stability (Allan deviation at 25±1C°, environmental effects are excluded) 1 s ≤5.0·10 −13
10 s ≤2.0·10 −13
100 s ≤7.0·10 −14
1 day ≤3.0·10 −15
Relative frequency accuracy ±3·10 –13 (factory calibration) ≤±1·10 –13 (in option with GPS calibrator)
Output signals frequency corrector resolution
tuning range
1·10 −15
1·10 −10
Phase noise spectral density (dBc/Hz) (10 MHz output) Frequency offset Spectral density (dBc/Hz)
10 Hz ≤−130
100 Hz ≤−150
1000 Hz ≤−160
Manual synchronization to external 1 pps TTL signal accuracy ≤±10 ns
Time synchronization to UTC using GPS ≤±20 ns (using optional GPS calibrator)
Temperature sensitivity in temperature operating range (10−40 C°) ≤±5·10 −15 1/C°
Magnetic sensitivity ≤±5·10 −15 1/Oersted

Full data monitoring and functions control. Application software running under Microsoft Windows.
Interface: RS-232C; LAN.
Power supply: AC(100÷240)V, (50÷60)Hz; DC(21÷72)V.
Power consumption: 100 V·A (AC), 100 W (DC).
Options:1. Built-in battery to prevent temporary power failure (li-ion 3 Ah); 2. Internal GPS calibrator for automatic calibration.
Dimensions (W×H×D): 483×133×560 mm.
Weight: 20 kg.
Warranty: 5 years.