Active Hydrogen Frequency Standard VCH-1005

The VCH-1005 is a high-performance hydrogen maser with a built-in frequency comparator. Extremely high frequency stability is provided by state-of-the-art technology. Ideal for high-accuracy measurement applications, particularly in metrology.

Key Applications
– national Time Keeping Service;
– deep space tracking and navigation;
– VLBI systems;
– scientific recearches.

Download manual

Specifications

Output signals:
– Sine: 5 MHz; 10 MHz; 100 MHz; (1±0.2) V RMS into 50 Ω load;
– Pulse: 1 Hz (pps), positive polarity, amplitude: >2.5 V into 50 Ω load; width: (15±5) µs; rise time:<10 ns.

Metrological characteristics are given in the table:

Relative frequency accuracy ±3×10 −13
(factory calibration)
Output signals frequency corrector resolution
returning range
1×10 −15
1×10 −10
Frequency stability (Allan deviation at (25±1)°C, environmental effects are excluded) 1 s ≤3.0×10 −13
10 s ≤3.0×10 −14
100 s ≤8.0×10 −15
1 hour ≤3.0×10 −15
1 day ≤3.0×10 −15
Long term frequency stability <1.0×10 -15 /day
Achieved in cavity autotuning mode together with another identical maser.
Phase noise spectral density (dBc/Hz)
(5 MHz output)
Frequency offset Spectral density
1 Hz ≤−115
10 Hz ≤−135
100 Hz ≤−145
1000 Hz ≤−155
Manual synchronization to external 1 pps TTL signal accuracy ≤±50 ns
Temperature sensitivity in operating temperature range (+10÷+35) °C ≤±1×10 −15 1/°C
Magnetic sensitivity ≤±2×10 −14 1/Oersted

Full data monitoring and functions control (local or remote through any network) performs by personal computer.
Software running under Microsoft Windows 2000, XP, Vista 7 is applied.
Power: 220 V, 50 Hz AC; or (22÷30) V DC.
Power consumption: 250 VA AC, 200 W DC.
Dimensions (W×H×D): 600×1585×900 mm.
Weight: 225 kg.
Warranty: 3 years.